The partial coherence method for assessment of impaired cerebral autoregulation using near-infrared spectroscopy: potential and limitations

D. De Smet*, J. Jacobs*, L. Ameye*, J. Vanderhaegen**, G. Naulaers**, P. Lemmers***, F. van Bel***, M. Wolf****, and S. Van Huffel*.

* Dept. of Electrical Engineering (ESAT), SCD Division, Katholieke Universiteit Leuven, Belgium.
** Dept. of Neonatology, University Hospital Gasthuisberg, Katholieke Universiteit Leuven, Belgium.
*** Dept. of Neonatology, Wilhelmina Children's Hospital, Utrecht, The Netherlands.
**** Clinic of Neonatology, University Hospital Zurich, Switzerland.

Submitted to the 2008 ISOTT Conference, Sapporo, Japan, Aug. 2008.

The most important forms of brain injury in premature infants are partly caused by disturbances in cerebral autoregulation. As cerebral intravascular oxygenation (HbD), regional cerebral oxygen saturation (rSO2), and cerebral tissue oxygenation (TOI) reflect cerebral blood flow (CBF), impaired autoregulation can be measured by studying the concordance between HbD/rSO2/TOI and the mean arterial blood pressure (MABP), assuming no changes in oxygen consumption, saturation (SaO2), and in blood volume. We investigated the performance of the partial coherence (PCOH) method, and compared it to the coherence method (COH). The PCOH method allows to eliminate in a linear way the influence of SaO2 on HbD/rSO2/TOI. We started from long-term recordings measured in the first days of life simultaneously in 30 infants from three medical centres. We then compared the COH and PCOH results with patient clinical characteristics and outcomes, and concluded that PCOH might be a better method for assessing impaired autoregulation.