Detection of Autoregulation in the Brain of Premature Infants
D. De Smet*, S. Van Huffel*, J. Vanderhaegen**, G. Naulaers**, E. Dempsey***
* Katholieke Universiteit Leuven, Department of electrical engineering, ESAT-SCD, Belgium
** Katholieke Universiteit Leuven, University Hospital Gasthuisberg, Belgium
*** McGill University Montreal, University Hospital, Canada
Submitted to the Belgian Day on Biomedical Engineering - IEEE/EMBS Benelux Symposium, Brussels, Belgium, Dec. 2006. Accepted for poster session.
Downloads
Poster available on-line at ftp://ftp.esat.kuleuven.be/sista/ddesmet/posters/0612.pdf
Conference paper available on-line at
ftp://ftp.esat.kuleuven.be/sista/ddesmet/reports/0612.pdf
Introduction
Brain injury in premature infants are caused in a
considerable part by disturbances in the cerebral
blood flow (CBF). Alterations in mean arterial blood
pressure (MABP) are very common in such infants
and otherwise the CBF can consequently often not
be maintained constant. Some studies [4] have
showed that an inadequate autoregulation of the
CBF occurs with concordant variations between
MABP and CBF. An important point is also that there
exists a strong correlation between the CBF and the
cerebral intravascular oxygenation (HbD), under the
condition that the arterial oxygen saturation (SaO2)
does not change appreciably. Several mathematical
algorithms were implemented to detect such
concordant variations between HbD, MABP and SaO2.
Methods
We have mainly used two ways to study the
concordance between some input signals : static
methods and dynamic ones. The static methods -
coherence (COH), correlation (COR), partial
coherence (PCOH) and principal autopower spectra
(PAPS) - can be applied between two (COH and
COR) or three (PCOH and PAPS) signals. The
dynamic methods, HTLS-SEP [1] and MUSCLE [2],
applied can process two (HTLS-SEP and MUSCLE)
or three (MUSCLE) signals simultaneously.
COH and COR are the more simple ways to study
the concordance between two signals. HTLS-SEP is
a measure to quantitate how similar the dynamics of
two signals are. It is derived from the common
subspace of the signals, and its goal is to fit one
linear model to two signals. MUSCLE is also derived
from the common subspace of the signals, but
makes use of semi-unitary matrices to determine the
number of common signal poles. The PCOH
measures the degree of linear dependence between
the spectra of two signals after having eliminated the
contribution of some other signals. PAPS is a
measure of the linear independence of the signals.
An important problem for the HTLS-SEP and
MUSCLE methods is the determination of the model
order of the input signals. For this purpose several
methods were studied, e.g. SAMOS [3] which is a
subspace-based method.
Experiments
HbD, MABP and SaO2 were measured at a frequency
of 4Hz on 6 premature babies, susceptible to suffer
from an impaired autoregulation, during a recording
time ranging from 8 to 16 hours for each. The
babies were aged from 24 to 28 weeks, and their
weights range from 675g to 1155g. The six methods
presented were implemented on these datasets.
Results and discussion
COR seems to be a stronger measure to detect
impaired autoregulation than COH. PCOH and PAPS
allow the consideration of a third signal. HTLS-SEP
and MUSCLE allow a dynamic analysis of the
measured signals, sometimes highlighting a dynamic
concordance between HbD and MABP even if COR
and COR are low.
References
[1]
Morren, G.; P. Lemmerling; S. Van Huffel; G.
Naulaers; H. Devlieger; and P Casaer.
Detection of autoregulation in the brain of
premature infants using a novel subspace-based
technique. Proceedings 23rd Annual
International Conference IEEE Engineering in
Medicine and Biology Society, Istanbul, Turkey.
Paper 4.5.1-7, 4 pages (CD-rom). October 25-
28, 2001.
[2]
Papy, J.-M.; L. De Lathauwer; and S. Van
Huffel. Common pole estimation in multichannel
exponential data modelling. Signal
Processing, vol. 86, pp. 846-858. 2006.
[3]
Papy, J.-M.; L. De LathauWer; and S. Van
Huffel, A shift invariance-based order selection
technique for the exponential data modeling.
IEEE Signal Processing Letters, 2006, to
appear.
[4]
Tsuji, M.; J. Saul; A. du Plessis; E. Eichenwald;
J. Sobh; R. Crocker;and J. Volpe. Cerebral
intravascular oxygenation correlates with mean
arterial pressure in critically ill premature infants.
Pediatrics, vol. 106, no. 4, pp. 625-632. October