• TITLE: Application of filtering methods for removal of resuscitation artifacts from human ECG signals
  • AUTHORS: I. Markovsky, Anton Amann, and S. Van Huffel
  • ABSTRACT: We consider the application of band-pass, Kalman, and finite impulse response adaptive filtering methods for removal of resuscitation artifacts from human ECG signals. Using a database of separately recorded ECG and resuscitation artifacts signals, we study to what extent the assumptions required by the methods are satisfied. Adaptive filters are based on a reference signal that is well correlated with the artifacts and is uncorrelated with the pure ECG signal. Such a reference signal is measurable in practice and is available in the database. The band-pass filter does not use a reference signal but is based on the assumption that the spectra of the artifacts and ECG signals are separated. This assumption is well satisfied for the signals in the considered database. The methods are evaluated on artificially constructed artifacts corrupted ECG signals. The considered performance criterion is the signal-to-noise (SNR) improvement, defined as the ratio of the SNRs of the estimated signal and the given ECG signal~$y$. Empirical results show that for low SNR of~$y$, the best performance is achieved by the band-pass filter, which is the simplest and computationally most efficient filtering method. For high SNR of~$y$, the best performance is achieved by the Kalman filter. However, a real-time implementation of the Kalman filter is an adaptive filter, which performance is worse than the one of the band-pass filter.
  • KEYWORDS: Kalman filter, adaptive filter, ECG signals, artifacts removal, system identification.
  • STATUS: Internal report.
  • DATE OF ENTRY: December 2006

  • paper: compressed ps file
  • software: compressed file

    The archive contains: ecg_plots.m - an m-file producing the plots in the paper, ecg_test.m - an m-file testing the methods, and ecg_lpf.m - a function implementing the low-pass filtering method.

    ecg_plots.m requires software for approximate linear system identification and ecg_test.m requires the System Identification toolbox of Matlab.


  • BibTeX entry:
  • @TechReport{MAV06,
      author = 	 {I. Markovsky and A. Amann and S. {Van Huffel}},
      title = 	 {Application of filtering methods for removal of resuscitation artifacts from human {ECG} signals},
      institution =  {Dept. EE, K.U.Leuven},
      year = 	 {2006},
      number = 	 {06--212}
    }